PRODUCT INTRODUCTION
The HI7061L General Purpose pH Electrode Cleaning Solution is a 500 mL formulation designed to remove mineral deposits and surface films from pH and ORP electrodes. Regular use of this cleaning solution prolongs electrode life and ensures measurement accuracy by restoring proper junction flow and offset voltage.
FEATURES
IMPORTANCE OF CLEANING
Dirty or improperly cleaned electrodes cause drift and inaccurate readings. During calibration, the meter compensates for any offset voltage; ideally this offset remains within ±30 mV. If calibration is performed with a fouled junction, subsequent measurements will retain that error, leading to unreliable pH or ORP values.
USAGE
PACKAGING & TRACEABILITY
SPECIFICATIONS
| Parameter | Detail |
|---|---|
| Product Code | HI7061L |
| Volume | 500 mL |
| Composition | Proprietary pH/ORP cleaning agents |
| Application Time | 15–20 minutes |
| Shelf Life (unopened) | 5 years from manufacture |
| Storage | 10 °C to 30 °C, out of direct sunlight |
| Packaging | Amber PET bottle with tamper-proof seal |